What is a Thermal Mass Flow Meter & What is its Working Principle? - Part 2 of 2

Posted by Doug Baker

Tue, Jun 09, 2015 @ 11:58 AM

This is part two of a two-part blog on Thermal Mass Flow sensors.  In part one, we described the desired characteristics of a thermal mass flow sensor.  In part two, we will discuss the operation of the 300 series flow sensor (Patent #6,125,695) and how its design addresses the desired traits.

300_series_flow_sensor_insideIn our previous blog, we showed a cutaway of a thermal mass flow meter.  Now let’s take an inside look at the 300 series flow sensor:

When gas is flowing through the bypass shunt, a small pressure drop is developed which will direct a fraction of the flow through the arced / semi-circular capillary tubing in the flow sensor. On the outside of the capillary tube, there are two resistive wire coils which are tightly wound and in excellent thermal contact with the tube. These two identical windings are referred to as:

  • Upstream Heater Coil (1)
  • Downstream Heater Coil (2)

Associated with each of the two heated coils is an ambient coil. The ambient coil is in excellent thermal contact with the aluminum ambient block.  Aluminum has a very high thermal conductivity which ensures that both ends of the sensor tube and the two ambient coils (3 and 4) will be at the same temperature.

heated_coils_upstream_downstreamTwo identical Wheatstone resistance bridges are formed from the two pair of coils (see image on right).

The circuit shown in the image on the right is designed to ensure that the heated coils (upstream and downstream) are maintained at a constant temperature (ΔT) above the corresponding ambient coils.

Next, we calculate the power (W) required to maintain ΔT by:

Power_Formula

This power will be calculated for both the upstream bridge and the downstream bridge. It can be shown that:

 Upstream_downstream_bridge_formula

So, by maintaining both heaters at the same ΔT above ambient, the mass flow rate is directly proportioned to the difference in power (W) between the two bridges. For example, when no flow is passing through the capillary sensor tube, the power needed to maintain ΔT will be the same (i.e. ṁ = 0)

As gas flow increases in the tube, heat is transferred from the upstream heater to the gas stream.  This will force the upstream circuit to use more power to maintain ΔT. In turn, the gas will transfer heat to the downstream heater which will cause the downstream circuit to use less power to maintain ΔT.

LinearityNow, here is the best part: the mass flow rate is directly proportional to the power difference. In other words, LINEARITY!

In our previous blog, we discussed how excellent linearity leads to improved accuracy. And, not only does the 300 series sensor give excellent linearity, the circuit shown on the right reacts very fast to changing flow. Thus, the 300 series has excellent responsive time.

One last note, we have designed the 300 series to use relatively large diameter tubing.  This larger tubing allows flow meters to be designed with lower pressure drop than many mass flow meters on the market.  

Visit our website for more information on Teledyne Hastings 300 series Flow Meters.